Pressure regulated volume control

Pressure regulated volume control (PRVC) is a ventilator mode where the breaths are delivered mandatorily to assure preset volumes, with a constant inspiratory pressure continuously adapting to the patient's condition. The flow pattern is decelerating. This mode is a form of assist control mode of ventilation, the breaths can either be ventilator initiated or patient initiated. This mode combines the advantages of volume controlled and pressure controlled ventilation.

How the ventilator achieves PRVC mode

The first breath delivered to the patient is a volume controlled breath. The measured plateau pressure is used as the pressure level for the next breath. The pressure is constant during the set inspiratory time and the flow is decelerating. The set tidal volume is achieved by automatic, breath-by-breath pressure regulation. The ventilator will adjust the inspiratory pressure control level, according to the mechanical properties of the airways/lung/thorax, to the lowest possible level to guarantee the preset tidal volume. If the measured tidal volume increases above the preset, the pressure level decreases in steps of maximum 3 cmH2O (300 Pa) between consecutive breaths until the preset tidal volume is delivered. Maximum available pressure level is 5 cmH2O (500 Pa) below a preset upper pressure limit.

References

External links

Wikipedia-logo.png The basis of this article is contributed from Wikipedia.Org. These articles are licensed under the GNU Free Documentation License It may have since been edited beyond all recognition. But we thank Wikipedia for allowing its use.
Please discuss further on the talk page.

  • Currently 0.00/5

Rating: 0.0/5 (0 votes cast) login to rate

Add to Favorite Print This Page Publish on Twitter
Bookmark and Share
close about Number of comments per page:
Time format: relative absolute
You need JavaScript enabled for viewing comments